Parker 910-000044-050 - 55 Series 10655

Thermally compensated 2-way normally closed (NC) proportional valve designed to maintain accurate and repeatable flow over a wide range of media. Ideal for medical and analytical applications that require precision control and flow up to 56 slpm.
Availability:
Partner Available:Quote Required
Brand Parker
Part No. 910-000044-050
TBD
Ship to
*
*
Shipping Method
Name
Estimated Delivery
Price
No shipping options
Products specifications
Attribute nameAttribute value
FunctionNormally Closed (NC)Seal MaterialFKM
For Fluid TypeGasesMaximum Operating Pressure150 psi
Configurationn/aMaximum Operating Temperature131 °F
Actuation TypeSolenoidDivisionPrecision Fluidics Division
Operation TypeElectronicMinimum Operating Temperature0 °C
Control MethodCurrent or VoltageMax Order Quantity9
Standard Lead Time20Out of Stock Lead Time40
Shipping Weight0.35Leak Rate< 0.016 cm3/min of Helium @ 150 Psi
Mounting TypeMounting ScrewOrifice Size0.76 mm
Length1.79 inchWeight2.2 oz
Body MaterialBrass 360 H02Width15.9 mm
Input Voltage29 VDCConnection Size4-40 x 5/8" Pan Head Phillips Screw
Cycles per Lifetime100 MillionThermal CompensationThermally Compensated
Electrical Connection4-Pin PCBHeight17.0 mm
Flow Rate23 SLPMConnection TypeManifold Mount with Screens
Maximum Flow Rate23 SLPMPower Consumption2.0 W Maximum
Industry ApplicationOxygen Conservers & ConcentratorsVacuum to Pressure Range686.0 mm-Hg to Pressure Rating
Body StyleManifold Mount w/ ScreensPneumatic ConnectionSquare Body(Manifold and Barb)
Electrical ConnectionPC Board MountBody/Elastomer MaterialsBrass/FKM
Maximum Operating Voltage29 VDCMaximum Operating Pressure/Orifice Size150 psid/.030"
Product resources